Packaging Leak Tester MFT-1000
Medicines: bottled, bagged, boxed, ampoules, vials, cartridges, prefilled needles (PFS), (BFS), (FFS), etc.
Spray cans: plunger spray cans; bag-lined spray cans; "energy jacket" spray cans; flexible tube spray cans.
Comply with USP <1207>, ASTM F2338 standards and FDA standards
Non-destructive non-destructive testing, high accuracy, repeatability, sensitivity
The leak rate can be automatically converted into defect aperture µm
Database storage of test results for easy quality management
Semi-automatic detection, suitable for small batch and multi-variety testing
The instrument is used for vacuum pressure, pressure decay difference detection
Database storage of test results for easy quality management
Touch-type man-machine interface, simple and quick operation: after setting/selecting the test program, only need to manually put in/take out the test sample
Adpot automous optimization Linux system
Automatic leak flow calibration function
With audit trail function
Split design, the test chamber is located above the host, and various test chambers can be provided according to different product types
The test cavity is customized according to customer needs to ensure that the test cavity is fully matched with the customer’s product, and rapid and sensitive testing
Automatically test the flow rate and change the aperture size during the whole process
Equipped with standard leaks (standard leak bottles, with third-party certification)
Four-level user authority management meets FDA 21CFR PART 11 requirements
We also provide users with supporting services related to tightness testing, including positive bottle(mould) production, standard leak rate/annual leak verification, new sample mold customization, sample methodological parameter development and verification, etc
| Differential Pressure Sensor | ±2kpa, error≤0.5%F.S. | |||
| Differential Pressure Resolution | 0.1pa | |||
| Test Pressure Sensor | Error≤±1%F.S. | |||
| Test Pressure Range | -0.1~0.2Mpa /-0.1~0.7MPa/-0.1~ 1.0Mpa (operational) | |||
| Minimum Detectable Aperture | 2µm (for IV bag minimum 5µm) | |||
| Display Method | 10-inch touch color touch screen | |||
| Voltage | AC100V-240V, 50/60Hz, power ≤200W | |||
| Weight | Host: about 30kg | |||
| Audit Trail | >=5 years of storage Multiple event logs, which can be queried according to time | |||
| Authority Management | Login with username and password, level 4 authority, meet FDA 21CFR PART 11 requirements | |||
| Print Function | Built in printer | |||
| History Record | >=300000 storage | |||
| Data Backup | Support U disk to export data | |||
| Interface | Ethernet USB RS485 (operational) | |||