Product Features
Calibration interface
Offer temp.and pressure interface,together with the 3Q adapter joints,can insert max 15 sensors one time to do the calibration work
Agar program
Can heat up, melt, warm the agar or used as chamber preheating function
Program Memory storage system
User can set their own program,and saved in system
Waste program
Special waste program to sterilize the waste thoroughly
Option forall model
Special waste program to sterilize the waste thoroughly
Saturated steam monitoring
The system automatically monitors the purge out of cold air to ensure the pure steam environment
Air filter (only for SR models)
Filtration is higher than 99.99%in particle with diameter of 0.2µm for air inlet
SUS316L Stainless steel (Option)
Stronger corrosion, longer service life, and greater pressure resistance
Auto start up appointment
Can make an appointment in 0-15 days
Smart microcomputer control
Inspiration”high speed microcomputer control,not only to monitor and control the whole process,but also to update the function in the future
LCD/USB/APP/Smart management system (Option)
LCD: 5 inch LCD screen to show information in words
USB: to help to store and transfer the sterilization data
APP: Remote management APP (only in China)
Smart management system: linked with
LIMIS/SCADA system(only in China)
Condenser (Option)
Equipped with a professional condenser,the steam generated during the sterilization process is converted into water through the, achieving zero discharge of water vapor/mist
Safety Features
Humanized design
The lid and bench corners are covered with insulation to avoid getting scald
The exclusive patented “eight column evenly distributed”interlock device is safer and more
Interlocking device
Reliable than ordinary interlocking devices
Lid closing checking
Machine can not start if lid is not locking well by checking mechanically and electrically
Real time monitor
System will monitor the lid locking,pressure and temperature status on a real time basis to guarantee the safe and successful work of the autoclave
Cooling lock OPEN temperature
Each program can set different lid OPEN temperatures,especially for liquid program ,so as to prevent userfrom opening lid and causing burns when the articles are still in a high temperature state
Triple protection for lack of water
The bottom of the sterilization chamber is equipped with three different dry burning protection devices:liquid expansion type,copper temperature sensing type,and electric ion type(water level sensor),to avoid misjudgment caused by a single method
Double protection for over pressure
Mechanical pressure gauge and electric sensing system,once pressure is abnormal,system will cut power,release the pressure and send out alarm
Auto trouble shooting
System will monitor the working status,and willcut the power,send out alarm,and report the error information once any abnormal situation detected
Safety test function
Regular testing of system security and safety protection measures is possible
Electrical protection
Equipped with a highly sensitive electrical protection system,effectively providing leakage protection, overcurrent and short-circuit protection
Spare Part Option

Safety Features
| 模型 | Standard GR60SA | Standard GR85SA | Standard GR110SA | |
| Auto Water Ill GR60SF | Auto Waterfill GR85SF | Auto Waterfill GR110SF | ||
| Auto Drying GR60SR | Auto Drying GR85SR | Auto Dying GR110SR | ||
| Capacity (L) | 60 | 85 | 110 | |
| Dimension (L*W*H)(mm) | SA:644x660x980 | SA:64x660x980 | SA:644x660x980 | |
| SF/SR:644x831x980 | SF/SR:644x831x980 | SF/SR:644×831×1180 | ||
| Chamber Dimension (Dia*H)(mm) | ①390×505 | ①390×700 | ①390×895 | |
| Rated Power(W) | 2900 | 4600(2900 as option) | 4600 | |
| SUS 304 Basket (Dia*H)(mm)(Note 1) | (370×190)×2 | ( φ370×280)×2 | ( φ370×250)x3 | |
| Power Requirement | 220V±10%15A 50HZ/60HZ | 220V±10%25A 50HZ/60HZ | 220V±10%25A 50HZ/60HZ | |
| Temperature Display Acuracy | 0.1°C | |||
| Working Environment | 5°C~40°C relative humidity≤85% | |||
| Chamber Material | SUS304 stainless steel | |||
| Sterilization Temp.range | 105°C~138°C | |||
| Sterilization Time.range | 1~6000 min | |||
| Melting Temp.range | 60°C~115°C | |||
| Melting Time | 1~6000 min | |||
| Warming Temp | 45°C~79°C | |||
| Warming Time | 1~9999 min | |||
| Drying (Only for SR) | Drying time: 1-300min Drying temp: 80-160°C | |||
| Cooling lock OPEN Temp.range | 40°C~99°C for solid,waste,agar,self-defined 40°C~80°C (note 2)for liquid | |||
| Sterilization Mode | Liquid, Liquid with warming, solid, Wrapped instrument, Fabric, Ruber, Fast, Waste, Agar, Self-defined | |||
| Program Number Saved | SA:20 programs SF/SR:60 programs | |||
| Exhaust Speed Control | Fully automatic internal discharging with 6 levels choice | |||
| Lid Open Way | Flip open (up and down) | |||
| Water Tank | SA: 2 water tanks to collect steam SF: 1 water storage tank more than SA SR: I water tank to collect steam+1 water storage tank (note 3) | |||
| Appointment Timer Range | (Auto start up appointment) 0~15 days | |||
| Pressure Gauge Range | -0.1-0.5MPa | |||
| Pressure Valve Set Pressure | 0.29Mpa | |||
| Controler | “Inspiration”smart microcomputer controller | |||
| Safety Device | “Eight column evenly distributed” interlock device, electric double inter lock, lid closing checking,dry scorch protection, safety valve, water level sensor,over temperature/pressure protection, temperature/pressure monitor, electricity safety (over current/short circuit/leakage), cooling lock, anti-scald lid and bench corner, automatic troubleshooting | |||
| Biosafety Function | Option for0.2um biosafety filter (basic model+S) | |||